Technical Data
Operating Mode: Intermittent opeation 30s on, 9 min off
Triggering of The Light Source (LED)
Nominal Voltage of The KaVo Mini
Led Lamp: 3.4V DC
Maximal LED Current: 150 mA
Voltage Range of The KaVo Mini LED Lamp: 3.0 – 3.6V DC
Media Supply at Motor Intake
Cooling Air (System Pressure): Max. 5.0 bar (73 psi)
Amount of Cooling Air at Motor Intake: 20 ± 2 NL/min
Spray Air: 1.0 – 2.5 bar (15 – 36 psi)
Spray Water: 0.8 – 2.0 bar (12 – 29 psi)
Media Exit at Instrument Coupling
Cooling Air Exit at Instrument Coupling :7.5 ± 1.5 NL/min
Spray Air: 3.5 + 2 NL/min at 1.0 bar (15 psi)
Spray Water: > 75 ml/min at 0.8 bar (12 psi)
Air Requirements
Air Quality in Accordance with ISO 7494-2: dry, oil-free, dirt-free, non-contaminated
Air Filter, Supplied by Customer: <20µm
Water Requirements
Water Quality in Accordance with ISO 7494-2: Tap Water
pH: 7.2 – 7.8
Water Filter, Supplied by Customer: <80µm
Opeating Conditions
Temperature: +10°C to +35°C (50°F to 95°F)
Maximal rel. Humidity: <80%
Air Pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)